  Buy This Autosampler For:
 
  • It can do everything!
  • Fixed or Variable volume
  • Heating and chilling
  • Dilution
  • Sample derivitisation
  • Preparative HPLC
 
     
Features & Benefits   Specification
  • Choice of Fixed or Variable volume mode.

    Use fixed volume for:
    i) lower cost
    ii) QC type applications
    iii) highest injection reproducibility.

    Use variable volume for flexibility and research applications
  Sample tray:

2 x 60, 1.5ml vials

 
or 2 x 24, 5ml vials
Wash bottle:

500ml

Std syringe:

500ul

Injection per vial:

1 - 91

Analysis time:

Up to 999 mins

   
Syringe Stepper Motor:-  
Syringe(ul):

Resolution:

  • Heating and chilling option. Use heating to keep samples in solution, and chilling to prevent thermal degradation
100

 175 steps/ul

250

 70 steps/ul

500

 35 steps/ul

  • Sample pre-column derivatisation option. Up to three reagents can be used, and reacted with the sample with independent reaction times. In addition the sample can be diluted up to 100 times
  1000

 15 steps/ul

2500

 6 steps/ul

5000

 3 steps/ul

   
Reproducibility:

< 1%, (10ul variable vol) 

   
<0.5% Fixed Volume
  • A totally metal free version is available
Memory Effect: 

< 0.01% 

  • Preparative version - Uses larger (5ml) syringe, wider bore needle and tubing, rack for 2 x 24, 5ml vials
     
Dimensions(mm)
300(h) x 300(w) x 400(d)
Weight 

24kg

  • Syringe drive in 17000 steps, so very reproducible injections are possible
Power

240V 50Hz

   
  • Up to 10 wash programs ensure the needle is clean so sample carryover is minimised. Washing solvent is isolated from sample by an air segment to ensure the most accurate injection
     
   
   
   
  • Large display shows all active information as a single display
   


Ordering Information 
Autosampler, fixed volume S5200

10 50 001

Variable volume upgrade

10 51 001

Sample preparation upgraded

10 51 002

Inert version (PEEK)

10 51 004

Preparative version

10 51 005